ULTRACO2 Laser
Self-diagnosis system for safety of CO2 LASER

 |
Advantage of HM – 30 |
 |
|
-
Thermal effect has been maximized geared to moisture content,
blood flow, heat transfer of mutually different biological
organism.
-
The
strength of penetration has been reinforced by stabilizing the
laser wave within visual scope.
-
It
instantly changes thermal effect of Ultra super pulse laser of
human body.
-
It
can select either from continuous or super pulse or ultra super
pulse mode.
-
With
the provision of system self-diagnosis sensor condition of the
machine can be checked very simply.
-
Cooling water automatic sensing function (When the cooling water
is insufficient it breaks off power.)
-
At
the time of occurrence of abnormal power it is possible to stop
the operation.
-
Convenient control panel (System operation command message) so
any user can use it without any difficulty.
-
Diverse hand piece and accessories treatment of various wounded
parts is possible..
|
|
 |
Application |
 |
|
-
Skin N plastic
surgery
-
Spot, freckle, age spot, acne, small wrinkles.
|
|
 |
Technical Specification |
 |
| Laser Type |
CO2 (Sealed off) |
| Wavelength |
10,600 nm |
| Mode Structure |
True TEM00 Mode |
| Power to Tissue |
Continuous and pulsed: 1W to 30 W Super Pulsed(Char-Free): 0.1 W to 9.9 W Ultra Pulsed (Char-Free): 0.1 W to 3 W |
| Pulse Duration |
Super Pulse: 400-800us, Ultra Pulsed: 300us less then |
| Peak Powder |
Up to 300-600W |
| Hand Pieces |
50 mm and 100mm,
ENT(optional) |
| Spot size |
100mm Hand Piece: 1.0mm, 50mm Hand Piece: 0.2mm |
| Display |
Graphic LCD or FND |
| Memory |
User programmable settings |
| Beam Delivery |
Articulated arm with 7 mirrors, hollow fiber (option) |
| Aiming Beam |
635nm, 5mW, adjustable brightness, blinking, and off while lasing |
| Cooling System |
Closed cycle water to air |
| Power Requirement |
3/6A, 110/220V, 50/60Hz, Single Phase |
| Dimensions |
350mm(W)x300mm(l)x1110mm(H)(without arm) |
| Weight |
40kg |
|
|
|
|