JoRak the system of
choice for Targeted Phototherapy.

-
JoRak provides both
UVA & UVB light within one system for increased treatment
benefit.
-
Simple system
interface for UVA and UVB settings.
-
MED & MPD Auto
Phototest Mode.
-
Simple user
controls.
-
JoRak can be
operated by staff nurse and phototherapists.
-
In-built calibrating
port for easy automated calibrating procedure.
-
Square aperture
handpiece design provides clear view.
-
Lightweight for easy
handling.
-
Flexible liquid
lightguide for easy access to areas which are difficult to
reach.
-
Disposable tip for
hygienic single patient use prevents cross contamination.
-
Transparent material
for clear visualization.
-
Uniform energy
distribution to skin.
-
No hot spots
-
Large 20 X 20 mm
square aperture.
-
More patients
treated in less time which in turn maximizing their return on
investment.
|
|
 |
Indications |
 |
|

Psoriasis, Vitiligo,
Atopic Dermatitis etc., |
|
 |
Advantages |
 |
-
Low cost & High
Performance
-
High dosing
capability affords quicker treatment time
-
Clearing of
psoriasis plaques in 4 – 6 treatment sessions
-
Pleasant treatment
experience
-
Long term
reliability
-
Compact system with
an optional Trolley
-
Can be conveniently
transported to multiple locations
|
|
 |
Principle
of Microdermabrasion: |
 |
|

A small hand piece
directs a stream of very fine crystals onto the targeted area,
loosening the layer on dead skin cells that forms on the surface.) |
|
 |
Technical Specifications |
 |
Performance Specifications
|
UVA Wavelength
Range |
- |
330 – 380 nm |
|
UVB Wavelength
Range |
- |
290 – 330 nm |
|
UVA Intensity (Handpiece) |
- |
100 – 550 mW/cm2
for 4cm2 exit aperture |
|
UVB Intensity (Handpiece) |
- |
50 – 250 mW/cm3
for 4 cm2 exit aperture |
|
User selectable |
- |
UVA Doses 0.4 –
5.0 J/cm2 |
|
User selectable |
- |
UVB Doses 15-3120
mJ/cm2 |
|
Calibration range |
- |
50 – 500 mW/cm2+/-20%
output |
|
Dose Accuracy |
- |
+/-20% or +3.3 mJ/cm2
of displayed value,
(whichever is greater) |
System Specifications
|
AC Supply
Connection |
- |
240 V, 50/60Hz |
|
Equipment Class |
- |
I |
|
Operation |
- |
Continuous |
|
Dimensions (LXBXH) |
- |
35cmX47cmX16cm |
|
Weight |
- |
11.5kg |
Operating Environment
|
Temperature |
- |
10ºC to 40ºC |
|
Relative Humidity |
- |
30% to 75% |
|
Atmospheric
Pressure |
- |
700 hPa to
1060 hPa |
|
|
|